Electrical Connections - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS031.07E
8

Electrical connections

Installation guidelines
IMPORTANT:
A ground fault interrupter (GFI) (I N = 30 mA) must be installed at a suitable point in
the circuit (in compliance with VDE 664).
Mains connection in the
installation room
Mains cable
(part of the delivery)
Network cable
(part of the delivery)
Main switch
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 1
12-2007
Electrical installations have to be done in accordance
with national regulations. Examples:
(Germany) Electrical installations in the
VDE
installation room must be in compliance with the
regulations IEC 364, VDE 0100, and VDE 0107.
(US) Electrical installations in the installation
UL
room must be in compliance with the regulations:
"National Electrical Code" (NEC) (NFPA70).
Double earthed pin outlet in compliance with
VDE
DIN 49441 and with CEE 7 standard cover V II.
Earthed contact outlet for the NEMA 5-15 P plug
UL
Euro-
Version:
US/CAN-
Version:
NEMA 5-15P
UTP; AWG24; Length: 5 m
Upon machine installation, it must be ensured that either
the mains connector or an all-pole circuit breaker for the
installation on site is located close to the machine and
easily accessible.
Printed copies of this document are not controlled.
DRYSTAR AXYS
5367/100
Installation Planning
CEE7/7 - H05VV - F3G - 3 x 2.5 mm² with
IEC C13 plug (female); cable length 3 m
NEMA 5-15P - SJT, 3 x AWG 18 with
IEC C13 plug (female); Cable length 3 m
US
IEC C13
IEC C13
Figure 8
Agfa Company Confidential
1
2
3
Europe
CEE7/7
5366_chap11_001.cdr
Chapter 11 / 21