AGFA DRYSTAR AXYS Service Manual page 196

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
Cure3
Part
HC305009 : thmodule.thermalheadheating
Type
Context
Action
Reason
Cure
Part
HC305010 : thmodule.voltage.low
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
Reason3
Cure3
Part
IMPORTANT:
This fault may have following additional reason and cure:
Reason: Film Calibration files corrupted.
Cure: Reset the calibration data for the film calibration of the upper and lower tray,
followed by a film calibration for both trays.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Replace
OSB
Failure
W305009
SeeManual
TH module heater defect
Replace
MOD05
Failure
W305010
SeeManual
Serial connection to powersupply fails
power down the printer, and power up again
PSU module defect
Replace
MOD09
TH module defect
Replace
MOD05
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting - Error Codes