DD+DIS033.07E
1

Safety Notes for Service Actvities

1.1

Safety Notes for Installation Planning and Installation

CAUTION:
The device is heavy (90 kg / 198 lb). Risk of injuries.
At least three people have to lift the device from one position to the other.
Use proper foot and hand protection when lifting the device.
The table might break and cause injuries or damage to the device. The device weight
may be too heavy for the table.
Communicate to the customer, that the table must be able to carry more than 90 kg
(198 lb).
1.2

Safety Notes for preventive and corrective Maintenance Work

WARNING:
Thermal Print Head is hot - up to 70 °C: Risk of burns.
Avoid contact with the Thermal Print Head.
Observe the relevant sticker on the Thermal Print Head.
CAUTION:
Thermal Print Head is fragile: Risk of damage.
When replacing or cleaning the Thermal Print Head great care must be taken.
CAUTION:
When mounting the main drive belt in the wrong direction, the image will be
corrupted:
To ensure correct reassembly mark the outside with a waterproof pen before removing
the belt .
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
09-2007
Safety Guidelines / General Repair Instruction
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367
Repair and Service
Chapter 3.1 / 4
Agfa Company Confidential