DD+DIS033.07E
Internal Densitometer Calibration
NOTE:
Always perform a Film calibration (refer to 2.1) and check the Density reproduction
(refer to 3.1) after a CDM calibration.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Welcome to this
DryStar Axys printer
Service - Calibration
1.
Film Calibration
2.
Internal Densitometer Calibration
3.
Print Head Calibration
4.
Horizontal Alignment
5.
Smart Profile Calibration
Go back
Welcome to this
DryStar Axys printer
Service - Calibration
Upper Tray
Lower Tray
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Adjustments and Calibrations
Service - Calibration
Internal Densitometer Calibration
Select Densitometer
Start Calibration
Enter below the measurements of the printer calibration sheet
Step Previous values
00
02
04
06
536736ak.cdr
08
10
12
14
16
18
Accept changes
Figure 3
Type 5367/100
Repair and Service
Welcome to this
DryStar Axys printer
Macbeth TR924
Gretag
X-Rite 301
X-Rite 310
Go back
Reset Values
Welcome to this
DryStar Axys printer
Internal Densitometer Calibration
SUCCEEDED
Step Previous values
0.24
0.25
01
0.25
0.25
0.26
03
0.30
0.55
0.54
05
0.94
1.45
1.43
07
1.96
2.49
2.50
09
3.00
3.15
3.18
11
2.80
2.28
2.28
13
1.71
1.20
1.17
15
0.71
0.38
0.37
17
0.26
0.24
0.25
19
0.25
Discard changes
Agfa Company Confidential
0.24
0.31
0.95
1.97
2.99
2.80
1.75
0.72
0.26
0.24
Reset values
Chapter 3.6 / 11