DD+DIS033.07E
3.7

ADVP

NOTE:
The latest version of ADVP can be downloaded from MEDNET GSO Library
(Hardcopy - Wet Imagers - LR DICOM Controller - Freeware)
DICOM troubleshooting tool
Purpose
ADVP can be used for following purposes:
Usage
#
Purpose
1
Send test images from the service PC to the printer
via network
2
Record DICOM print sessions for further
troubleshooting.
3
Record DICOM print sessions for changing and re-
sending them to the printer.
NOTE:
ADVP can run on the service PC or any Windows PC in the network. The only pre-
requisite: The PC must be part of the network.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
10-2007
Figure 50
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Tools and Auxiliary Means
Details see...
3.7.1.
3.7.3.
536632eq.cdr
Agfa Company Confidential
3.7.2
Chapter 3.2 / 49