Electrical Check; Basic Protective Earth Test - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS032.07E
3.9

Electrical Check

The electrical check is split up in two steps:
Step
1
2
3.9.1

Basic Protective Earth Test

(1)
Use a multimeter that is capable to
measure a resistance < 1 Ohm
(2)
Put one probe at the ground
connector of the mains input
(3)
Put the other probe at a few metallic
parts, the customer could touch.
The electrical resistance has to be
≤ 0.2 Ohm. Refer to Figure 14.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
09-2007
Task
Perform the basic protective earth test
Perform the electrical check according to
National Regulations.
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367
Maintenance
Details see..
3.9.1
3.9.2
536631JN.cdr
2
Figure 14
Chapter 9 / 17