AGFA DRYSTAR AXYS Service Manual page 184

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
HC303024 : upperpickupmodule.s15
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC303025 : upperpickupmodule.s18
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
MODULE : lowerpickupmodule
Elabel : Yes
HC304000 : lowerpickupmodule.film.pickupposition
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Failure
W303024
None
Upper pickup pickup sensor S15 malfunctioning
check upper pickup pickup sensor S15
Upper pickup module defect
Replace
MOD03
Failure
W303025
None
Upper pickup passed sensor S18 malfunctioning
check upper pickup passed sensor S18
Upper pickup module defect
Replace
MOD03
Accidental
None
Lower input jam
Remove lower input jam
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 18 of 72