AGFA DRYSTAR AXYS Service Manual page 331

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
8
Software Release 2.1.0_C3
Type of new Software*
Release Date
Printer Serial Numbers
Availability
Prerequisites
*NOTE:
 A software upgrade is a functional extension. An update (software patch) is an
internal change designed to correct bugs, fix defects, or otherwise minor
improvements to performance without changing the software's basic design,
structure or functionality.
 On MedNet, GSO Library always the latest available software release can be
downloaded.
8.1
New Feature of Software 2.1.0_C3
By installing software 2.1.0_C3, the Devius board is re-programmed to revision 9.
This is a solution for the warning message "RF Tag Problem" after 10 printed films of a
new film pack.
For details see DRYSTAR AXYS SB 15 / DRYSTAR 5302 SB 20, Software 2.1.0_C3
available to solve Warning Message "RF Tag Problem" after 10 printed Films of a new
Film Pack, DD+DIS037.11E
NOTE:
Software 2.1.0_C3 supports DRYSTAR 5302 and DRYSTAR AXYS.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
03-2011
Software Releases and Patches
Software Update
02/2011
SN ≥ 3949 plus SN 3938
MedNet, GSO Library
None
(Intranet Link
/
Extranet
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Link).
Chapter 3.8 / 17