Print Head (Th) Calibration - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
2.3

Print Head (TH) Calibration

WHEN TO BE
PERFORMED
REQUIRED TOOLS:
INSTRUCTIONS
(1)
(2)
(3)
(4)
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
The Print Head Profile calibration is necessary whenever vertical stripes are not
longer acceptable:
Vertical stripes are unacceptable in case they are visible in the medical service
test image.
One of following, calibrated densitometers: McBeth TR-924, Gretag D200-2, X-
Rite 301, X-Rite 310, X-Rite 331, X-Rite 341
Lint free cloth
Isopropyl - alcohol
Print the Medical service test image.
Clean the thermal head with a cleaning tissue lightly moistened with Isopropyl -
alcohol. The last cleaning action must be carried out in one way.
In the webpages - menu 'Service Engineer Tools - Enter Service Command
session' select '6. Calibration - 3. Print Head Calibration'.
Follow the instructions on the WebPages. Refer to Figure 4.
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Chapter 3.6 / 12
Agfa Company Confidential