DD+DIS033.07E
HC311002 : inputsensormodule.connected
Type
Context
Action
Reason
Cure
HC311003 : inputsensormodule.s4
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC311004 : inputsensormodule.s13
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
E311002
SeeManual
XS5 not connected
Connect XS5
Failure
W311003
None
S4 not connected
Connect S4
S4 bracket module defect
Replace
MOD11
Failure
W311004
None
S13 not connected
Connect S13
S4 bracket module defect
Replace
MOD11