AGFA DRYSTAR AXYS Service Manual page 189

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
HC304016 : lowerpickupmodule.passed.high
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC304021 : lowerpickupmodule.m6.connection
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC304022 : lowerpickupmodule.m6.shortcircuit
Type
Context
Action
Reason
Cure
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Failure
W304016
None
Lower pickup past sensor jammed
Free lower pickup past sensor
Lower pickup past sensor defect
Replace
MOD04
Failure
W304021
SeeManual
Lower pickup motor not connected
connect lower pickup motor
Lower pickup module defect
Replace
MOD04
Defect
E304022
SeeManual
Lower pickup module defect
Replace
MOD04
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 23 of 72

Advertisement

Table of Contents
loading

Table of Contents