AGFA DRYSTAR AXYS Service Manual page 267

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
REQUIRED TOOLS:
• Calibrated densitometer MacBeth TR-924, X-Rite 331, 341, X-Rite 301, X-Rite
310 or Gretag D200-2
• Waterproof pen
REQUIRED TIME:
45 min.
(1) Perform a film calibration for this tray, where 14x17" films are available. See
section 2.1.
(2) Select in the browser menu
'Service Engineer Tools - Enter Service Command Session'.
(3) Select '6 Calibration - 5 Smart Profile Calibration'.
(4) Select the densitometer which will be used for measurement.
(5) Select the film tray with 14x17", 14x14" or 14x11" films: A test film as shown in
Figure 6 is printed.
(6) Measure the 66 squares on the printed film: Write down the values on the
foreseen squares on the film with a waterproof pen.
(7) Enter the values in the browser screen. See also Figure 7 and Figure 8.
IMPORTANT:
The browser screen where the 66 data are entered refreshes in a time interval of
4 minutes. If you are not finished with entering all data within 4 minutes, all entered
values will be reset to "0.00".
Workaround to avoid entering all data again: Press the "back button" in the browser.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Adjustments and Calibrations
Figure 6
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Chapter 3.6 / 15
Agfa Company Confidential