AGFA DRYSTAR AXYS Service Manual page 233

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
HC315002 : suspensiondrivemodule.s21
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC315003 : suspensiondrivemodule.s22
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC315004 : suspensiondrivemodule.connected
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
E315002
SeeManual
Sensor S21 not connected
Connect sensor S21
Suspension drive module defect
Replace
MOD15
Defect
E315003
SeeManual
Sensor S22 not connected
Connect sensor S22
Suspension drive module defect
Replace
MOD15
Defect
E315004
SeeManual
Connector XS13 not connected
Connect Connector XS13
Suspension drive module defect
Replace
MOD15
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 67 of 72