DD+DIS033.07E
3.4

Front Panel

Led
D1
D2..D25
D3
Color
Green
Red
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
10-2007
Electrical and Mechanical Codes, Fuses, LEDs
Color
green
green / red
green
Light
Status
Ready
Constant
(stand-by)
Startup
sequence
Receiving jobs
Calculating
Blinking
Printing
Calibrating
Installation
Key-operator
Service mode
Blinking
Film jam
Warning
Constant
Error status
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Function
12V power present
See table below
5V power present
Receive Jobs ?
No
No
Yes
Yes
Yes
Yes
No
Yes/No
No
Yes
Yes
No
Printing possible ?
No
No
Yes
Yes
Yes
No
No
Yes/No
No
No
Yes
No