AGFA DRYSTAR AXYS Service Manual page 92

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
Hardware Component
Key 6 Board
Hard disk
CF drive
Thermal head
OSB board
Power supply
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2009
Description
Controls the display with 122 (width) x 32
(height) pixels
Controls all buttons and LEDs required for
operation
Keeps the operating system, all application
software and diagnostic data
To save or install data via CF card type II
Transfers the diagnostic data (pixels) via
heating elements to the heat sensitive film
Switch board for the different TH heating
elements
Provides the power for heating up the resistor
elements
Supplies all voltages for the printer (+3,3 V,
+ 5 V, +12 V, -12 V, +20 V, +24 V, + 55V)
Connected via RS232 interface to the Devil
board, to allow saving temporary data before
switch off
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Functional Description
Agfa Company Confidentia
Chapter 2 / 22

Advertisement

Table of Contents
loading

Table of Contents