AGFA DRYSTAR AXYS Service Manual page 231

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
HC314005 : gearboxmodule.s3
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC314006 : gearboxmodule.connected
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC314020 : gearboxmodule.working
Type
Context
Action
Reason
Cure
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
E314005
SeeManual
Sensor S3 not connected
Connect sensor S3
Roller drive module defect
Replace
MOD14
Defect
E314006
SeeManual
Connector XS12 not connected
Connect Connector XS12
gearboxmodule drive module defect
Replace
MOD14
Defect
14
Replace
Roller drive module defect
Replace
MOD14
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 65 of 72

Advertisement

Table of Contents
loading

Table of Contents