Enclosed an example where it is required to configure two logical SCP printers:
Printing is initiated at a PACS workstation.
Two types of modalities as input.
o Mammography images: Interpolation cubic bell with
o CT images: Interpolation highres cubic with
Problem:
The PACS workstation has one AE title only. The DRYSTAR AXYS does not
recognize, for which type of modality (CT or Mammo) the user wants to print.
Solution (see "
Create one logical printer 'DSAXYSMammo' and one 'DSAXYSCT.
Create two different host profiles for the two logical printers with the required
settings.
In the PACS workstation setup both printers: The user has to select the proper
logical printer.
NOTE:
In the example of Figure 14
This setting prevents, that the modality can overrule the adjusted parameter.
It always depends on the application, whether this is desired or not.
This has to be discussed with the modality service engineer and/or the customer.