Film Calibration - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
NOTE:
It is also possible to enter the site data in
the "site data tool" on the service PC (see
MEDNET, GSO Library:
Hardcopy => Dry Imagers => DRYSTAR
AXYS => Freeware) and upload the so
created ZIP file via key-operator or service
menu "9 Installation - 3 Install site data".
4

Film Calibration

The purpose of this section is to show the tasks that are required for film calibration.
(1)
Load the films the customer will
use in the upper and lower film
tray. Refer to note below.
(2)
Select in key-operator menu
'6 Calibration - 1 Film Calibration'.
(3)
Select 'Start Calibration':
The calibration film as shown in
Figure 9
X
measured Dmax is shown.
See also note next page.
INPUT TRAY:
FILM CALIBRATION OK
Dmax = 3.20 (XR310)
Measured Dmax
(4)
Repeat film calibration for second
input tray.
NOTE:
The films delivered with the printer may only be used for test purposes.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Controls, Connections and Setup Procedure
is printed and the
X
536701aa.cdr
Default densitomete
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Figure 10
CDM measurement area
In order to be useable this film should be
absolutely dust free within the marked region!
Figure 11
Agfa Company Confidential
536536ga.cdr
Chapter 1 / 16