DD+DIS032.07E
4.2

Check of Safety Switch

(1)
Switch on the printer. Wait till it shows
the READY screen (takes approx. 7
minutes)
(2)
Open top cover
(3)
Check display message: It must show
"Close top cover"
4.3
SW Patch Installation
(1)
Install the latest SW patch, in case a recommended SW patch for the printer is
available. Refer to chapter 3.8, Software Releases, Patches.
4.4
Film Calibration
(1)
Select in key-operator menu
'6 Calibration - 1 Film Calibration'
(2)
Follow the instructions on the
display.
(3)
Repeat film calibration for second
input tray