8480 Series Power Sensor Specifications - Agilent Technologies 8481A Operating And Service Manual

8480 series coaxial power sensors
Table of Contents

Advertisement

NOTE

8480 Series Power Sensor Specifications

These specifications are valid with EPM and EPM-P Series of power meters.
The 8480 series thermocouple and diode power sensors provide accuracy, stability,
and SWR over a wide range of frequencies (100 kHz to 50 GHz) and power levels
(-70 dBm to +44 dBm)
Table 1-3
Frequency
100 kHz to 10 MHz
10 MHz to 30 MHz
30 MHz to 500 MHz
500 MHz to 1.2 GHz
1.2 GHz to 6 GHz
6 GHz to 14 GHz
14 GHz to 18 GHz
18 GHz to 26.5 GHz
26.5 GHz to 33 GHz
33 GHz to 34 GHz
34 GHz to 35 GHz
35 GHz to 40 GHz
40 GHz to 45 GHz
45 GHz to 50 GHz
* These uncertainties only apply to Option 033
Chapter 1
Typical measurement uncertainties of Calibration
Factor (CF) data printed on the power sensor
8481A
-
0.81
0.78
0.78
0.92
1.16
1.59
-
-
-
-
-
-
-
Introduction
25°C ± 3°C
8482A
8483A
8485A
0.87
1.59
-
0.8
1.39
-
0.8
1.41
1.29
0.8
1.41
1.26
0.9
1.46
1.35
-
-
1.61
-
-
1.77
-
-
2.47
-
-
3.35*
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
11