Calibration Factor And Reflection Coefficient - Agilent Technologies E Series Operating And Service Manual

Power sensors
Hide thumbs Also See for E Series:
Table of Contents

Advertisement

Specifications and Characteristics
E9300/1/4/A Power Sensor Specifications

Calibration Factor and Reflection Coefficient

Calibration Factor (CF) and Reflection Coefficient (Rho) data are provided
on a data sheet included with the power sensor. This data is unique to each
sensor. If you have more than one sensor, match the serial number on the
data sheet with the serial number on the power sensor you are using. The
CF corrects for the frequency response of the sensor. Agilent EPM series
power meters automatically read the CF data stored in the sensor and use it
to make the corrections.
Reflection Coefficient (Rho, or ρ) relates to the SWR according to the
following formula:
Maximum uncertainties of the CF data are listed in the following tables.
As the Agilent E- series E9300 power sensors have two independent
measurement paths (high and low power paths), there are two calibration
factor uncertainty tables for each sensor. The uncertainty analysis for the
calibration of the sensors was done in accordance with ISO Guide. The
uncertainty data reported on the calibration certificate is the expanded
uncertainty with a 95% confidence level and a coverage factor of 2.
Cal Factor Uncertainty
(Low Power Path,-60 to -10 dBm)
Frequency
9 kHz to 10 MHz
10 MHz to 30 MHz
30 MHz to 500 MHz
500 MHz to 1.2GHz
1.2 GHz to 6 GHz
6 GHz to 14 GHz
14 GHz to 18 GHz
42
ρ
1
+
SWR
=
------------ -
1 ρ
Uncertainty (25°C ±10°C)
E9300A
E9301A
-
-
±1.8%
±1.8%
±1.6%
±1.6%
±1.8%
±1.8%
±1.7%
±1.7%
±1.8%
-
±2.0%
-
Uncertainty (0°C to 55°C)
E9304A
E9300A
±1.7%
-
±1.7%
±2.2%
±1.7%
±2.0%
±1.7%
±2.5%
±1.7%
±2.0%
-
±2.0%
-
±2.2%
E9301A
E9304A
-
±2.0%
±2.2%
±2.0%
±2.0%
±2.0%
±2.5%
±2.0%
±2.0%
±2.0%
-
-
-
-