Ghz And 33 Ghz Frequency Operation; 8485D-033) - Agilent Technologies 8481A Operating And Service Manual

8480 series coaxial power sensors
Table of Contents

Advertisement

Table 1-2
Characteristic
11708A accuracy at
50 MHz, 25 °C
Dimensions
NOTE
NOTE
Introduction
11708A 30 dB attenuator characteristics
Limits
30 ± 0.05 dB
Length: 60 mm (2.4 in)
Diameter: 20 mm (0.8 in)
The 11708A 30 dB attenuator is intended for use only at the 1 mW, 50 MHz power
reference of the power meter. Its function as a calibration reference may be
compromised if used for other purposes.
26.5 GHz and 33 GHz Frequency operation (8485A, 8485A-033, 8485D and

8485D-033)

8480 series sensors that operate up to 26.5 GHz and 33 GHz are fitted with
APC-3.5mm (m) connectors as standard. To convert the APC-3.5mm (m) connector
for calibration an adapter (APC-3.5 (f) to Type-N (m)) is included with the power
sensors. Figure 1-4 shows the parts included with your power sensor.
The APC-3.5mm to Type-N adapter is intended for use only at the 1 mW, 50 MHz
power reference of the power meter. Its function as a calibration reference may be
compromised if used for other purposes.
8
Comments
Accuracy traceable to National Institute of
Standards and Technology (NIST), with a
temperature coefficient typically 0.003 dB per
°C.
Chapter 1