HEIDENHAIN ND 7000 Operating Instructions Manual page 283

Milling
Table of Contents

Advertisement

Servicing and maintenance | Updating the firmware
Updating the firmware
The firmware can be updated from a USB mass storage device (FAT32 format) or
via a network drive.
Safely removing a USB mass storage device
HEIDENHAIN | ND 7000 | Operating Instructions Milling | 01/2020
Tap Firmware update
Tap Select
If required, connect a USB mass storage device to a USB
port of the product
Navigate to the folder containing the new firmware
If you have accidentally tapped the wrong folder, you
can return to the previous folder.
Tap the file name that is displayed above the
list
Select the firmware
Tap Select to confirm the selection
The firmware version information is displayed
Tap OK to close the dialog
The firmware update cannot be canceled once the
data transfer has started.
Tap Start to start the update
The screen shows the progress of the update
Tap OK to confirm successful update
Tap Finish to terminate the service application
The service application is terminated
The main application is launched
If automatic user login is active, the user interface is
displayed in the Manual operation menu
If automatic user login is not active, the User login menu is
displayed
Tap File management in the main menu
Navigate to the list of storage locations
Tap Safely remove
The message The storage medium can be removed now.
appears
Disconnect the USB mass storage device
16
283

Advertisement

Table of Contents
loading

Table of Contents