CALCulate:MARKer Subsection

N9340B Programmer's Guide
Limit Line Testing
:CALCulate:LLINe[1]:[STATe] OFF|ON|0|1
:CALCulate:LLINe[1]:[STATe]?
Toggles the limit line testing between on and off.
If the limit line is turned on, limit pattern is automatically
turned off.
*RST:
Off
Key access:
LIMIT > Limit Line On Off
Limit Line Y-axis Value
:CALCulate:LLINe[1]:Y <ampl>
:CALCulate:LLINe[1]:Y?
Sets the Y- axis value of a limit line.
Limit line Y- axis value is set independently and is not
affected by the X- axis units.
*RST:
0 dBm
Key access:
LIMIT > Limit Line
Markers All Off on All Traces
:CALCulate:MARKer:AOFF
Turns off all markers on all the traces.
Key access:
Marker > All Off
Turn On 6 Makers on All Traces
:CALCulate:MARKer:ALL
Turns on all markers on all the traces and places them on
six highest peaks.
Key access:
Marker > Peak Search > 6 Peaks Search