Advertisement

5 Operation
104
EN
MEDUCORE Standard
7. Select new software with the navigation knob.
Damage to the device caused by moving the device and/or
pressing buttons during the update!
Moving the device and/or pressing buttons during the update may
cancel the update and damage the device.
⇒ Do not move the device.
⇒ Do not press any buttons on the device.
8. Update the device software with the Start function button.
9. Wait until the software has been updated.
After the end of the update, the device displays the message
SUCCESS.
10. Restart the device with the Restart function button.
The device restarts and the start menu appears on the display.
11. Call up the operator menu with the menu key
(see "8.1 Navigating the operator menu", page
12. Select Device information | Device information.
The device displays the installed software.
2
and enable
117).