Table of Contents

Advertisement

Risk of injury due to inoperational device!
Operation of the device after a failed function check may result in
injury to the patient.
⇒ Only operate the device after it passes the function check.
15. Proceed with the device according to the following table:
Display
Device is ready for use
Device is not ready for
use
Device is ready for use.
The service symbol
flashes in the start
menu.
16. End function check with the OK function button.
The start menu appears.
17. Disconnect the function test resistor from the trunk cable.
18. Connect the defibrillation electrodes to the trunk cable.
19. If necessary: Connect the ECG cable to the ECG cable
connection point.
20. If necessary: Connect the NIBP cuff to the NIBP connection
using the NIBP connecting tube.
Result
The function check is complete.

10.3 Checking the ECG cables

In addition to visual inspection of the ECG cables,
WEINMANN Emergency recommends checking function at regular
intervals
(see "10.2 Performing a function check", page
1. Connect the ECG simulator to the ECG cable connection using
the ECG cable.
Meaning
Function check passed
Function check failed
Function check passed,
but note about service
due
MEDUCORE Standard
10 Function check
Action
Use device without
restriction.
Contact your authorized
dealer or the
manufacturer.
Use device without
restriction, but have
device checked.
148):
2
EN
153