Advertisement

5 Operation
5.5.2 Manual defibrillation (only with Manual
Requirement
86
EN
MEDUCORE Standard
defibrillation option)
This function is only available if it has been enabled and activated
by the operator: Operator menu | System settings | Enable options
| Manual mode
(see "8.8 System settings", page
If you are the operator of the device and have access to the
operator menu, you can disable the manual mode: Operator menu
| System settings | Disable functions | Manual mode
System settings", page
The trunk cable and defibrillation electrodes are connected
"4.4 Connecting the trunk cable and defibrillation electrodes",
page
57).
The device is switched on
page
76).
The patient group is selected
group", page
78).
Risk of injury due to lack of knowledge and failure to follow
guidelines in the manual mode!
The use of the manual mode by users without medical
qualifications and training in defibrillation and/or failure to follow
guidelines can result in injury to the patient, user or bystanders.
⇒ Only use the manual mode if the user has a medical
qualification and is familiar with the operation of the device.
⇒ Follow the defibrillation guidelines.
⇒ Observe national and regional provisions on defibrillation.
⇒ Observe organizational guidelines on defibrillation.
Risk of injury due to missing battery in the AED mode and in
manual mode!
Without a battery, the capacitor for shock energy in the device
cannot charge. This prevents defibrillation and delays treatment.
⇒ Insert a battery before using the AED mode or the manual
mode.
2
137).
(see "5.1 Switching the device on",
(see "5.4 Selecting the patient
137).
(see "8.8
(see