10 Function check
Result
154
EN
MEDUCORE Standard
In principle, any ECG simulator can be used.
WEINMANN Emergency recommends the ECG simulator
WM 45444. Alternatively, the ECG cable can be tested on a
voluntary test person.
2. Switch on the ECG simulator and set a sinus rhythm.
3. Switch on the device
76).
4. Select the Adult patient group
group", page
78).
5. If necessary: Press the Monitor function button.
The device switches to monitor mode.
6. If the ECG curve is shown in the middle curve field: Press the
Lead button to show all ECG leads.
7. Assess the ECG leads:
All ECG leads must be shown.
All ECG leads must show the set sinus rhythm.
Shaking the cable must not result in one or more ECG leads
either not being shown or being shown with faults.
8. If ECG leads are not shown or are shown with faults: Replace
ECG cable.
9. Switch off the device
77).
10. Disconnect the ECG simulator and ECG cable from the device.
The ECG cable has been checked.