8 Operator menu
Parameter
Heart rate
Heart rate
Pulse rate
Pulse rate
SpO
SpO
Infant alarm limit
settings
NIBP systolic
NIBP systolic
NIBP diastolic
NIBP diastolic
Pause audio
Alarm tone
output in manual
mode
126
EN
MEDUCORE Standard
Possible
values
35/min-250/min,
in increments of 5
30/min-245/min,
in increments of 5
35/min-250/min,
in increments of 5
30/min-245/min,
in increments of 5
saturation
2
66 %-100 %
saturation
2
65 %-99 %
45 mmHg-
260 mmHg, in
increments of 5
40 mmHg-
255 mmHg, in
increments of 5
25 mmHg-
200 mmHg, in
increments of 5
20 mmHg-
195 mmHg, in
increments of 5
Infinite
1 min
2 min
5 min
10 min
Activated
Deactivated
2
Description
Here you can set the upper
( ) and lower ( ) limits for
the patient group as of which
the device must emit an
alarm.
According to the values set,
the setting ranges of the
upper and lower limits
depend on each other as
follows:
The setting range for the
upper alarm limit ends at
least one setting value above
the set value for the lower
alarm limit.
The setting range for the
lower alarm limit ends at
least one setting value below
the set value for the upper
alarm limit.
Here you can set the length
of time for which the audio
alarm output is paused. If you
select Infinite, the audio
alarm output is permanently
paused (the audio alarm
output is muted).
Here you can set whether the
audio alarm output is active
or inactive when the user
calls up the manual mode.
Factory
setting
200/min
100/min
200/min
100/min
95 %
85 %
100 mmHg
50 mmHg
70 mmHg
30 mmHg
2 min
Deactivated