10.4 Checking the NIBP cuff and NIBP
connecting tube
In addition to the visual inspection of the NIBP cuff and NIBP
connecting tube, WEINMANN Emergency recommends checking
function at regular intervals
check", page
To check the functionality of the NIBP cuff and NIBP connecting
tube, in addition to performing a visual inspection
Performing a function check", page
test:
1. Measure the blood pressure on a voluntary test person
"5.8 Performing non-invasive blood pressure measurement
(NIBP)", page
2. Watch for escaping air during the measurement to ensure that
the system is free from leaks.
3. If the NIBP connecting tube leaks: Replace the NIBP connecting
tube.
4. If the NIBP cuff leaks: Replace the NIBP cuff.
5. If the measurement takes too long or produces implausible
results: Check whether the patient group and NIBP cuff have
been correctly selected for the test person and that the NIBP
cuff has been attached correctly
cuff", page
6. Repeat the measurement.
7. If the measurement again takes too long or produces incorrect
results: Contact your authorized dealer or
WEINMANN Emergency.
Result
The NIBP cuff and NIBP connecting tube have been checked.