Non-Invasive Blood Pressure (Nibp) Monitoring - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Specification
Reference methods for determining the accuracy of
the pulse rate
Wavelength with maximum intensity
Curve form
Update rate of SpO
mean
2
Averaging
Delay in data updating
Delay in alarm condition
Delay in alarm generation
Functional oxygen saturation
Biocompatibility
15.9 Non-invasive blood pressure (NIBP)
Specification
Measuring method
Measurement range (systolic):
Adults
Children
Infants
Measurement range (diastolic):
Adults
Children
Infants
Accuracy (during operation between 0 °C and 50 °C) ±3 mmHg
monitoring
Pulse oximetry
Oxitest simulator testing
660 nm/890 nm (2.5 to 4.5 mW)
Normalized
8 s (It can take up to 16 s for the correct SpO
to be displayed).
12 s
500 ms
20 s
< 1 s
The pulse oximetry sensor is calibrated to display the
functional oxygen saturation and must not be
calibrated with a function tester.
The pulse oximetry sensor is latex-free. No
material used in its production contains latex
protein. The materials with which the patient
comes into contact have undergone extensive
biocompatibility tests. Further information is
available on request.
As per EN ISO 10993-10
Non-invasive blood pressure
measurement
Oscillometric, diastolic values correspond to phase 5
Korotkoff sounds
40 mmHg to 260 mmHg
40 mmHg to 230 mmHg
40 mmHg to 130 mmHg
20 mmHg to 200 mmHg
20 mmHg to 160 mmHg
20 mmHg to 100 mmHg
MEDUCORE Standard
15 Technical data
value
2
2
EN
183