10 Function check
10 Function check
10.1 Intervals
10.2 Performing a function check
Requirement
148
EN
MEDUCORE Standard
After being switched on, the device performs an automatic
function check which serves to check the functionality of all the key
functions.
The device also offers a step-by-step guide to performing a
function check
(see "10.2 Performing a function check", page
148).
When the device is switched on, permanent test routines are active
in the background which alert the user to malfunctions via an
alarm.
The correct functioning of some components cannot be checked
with the support of software. In such cases, the user must assess
functionality himself (e.g. external damage to the device,
functionality of the ECG cable or the NIBP cuff).
Part concerned
Device including accessories
•
The device is disconnected from the patient.
•
A battery with a battery status of at least 2 bars is inserted in
the device. If the battery status is less, the function check
cannot be started.
•
An SD card is in the SD card slot.
•
A pulse oximetry sensor connecting cable with pulse oximetry
sensor is connected.
1. Check the following parts for external damage:
•
Device
2
Interval
•
Before each use
•
After each hygienic reprocessing
•
After each repair