11 Alarms and faults
Alarm
NIBP diastolic
Settings lost
Check seating of ECG electrodes
ECG module defective
Plug in ECG connector
Heart rate
Heart rate > 250/min
Heart rate
Heart rate < 30/min
NIBP outside measuring range
No NIBP signal
160
EN
MEDUCORE Standard
Cause
Measured diastolic blood pressure
is below the set lower alarm limit.
Settings had to be reset to factory
settings.
ECG electrodes not attached or not
correctly attached to the patient or
the ECG cable is not correctly
connected to the ECG electrodes.
Internal ECG module defective.
ECG connector of the ECG cable
not connected or incorrectly
connected to the device.
Measured heart rate is above the
set upper alarm limit.
Measured heart rate is above
250/min (device shows --- as heart
rate).
Measured heart rate is below the
set lower alarm limit.
Measured heart rate is below
30/min (device shows --- as heart
rate).
Blood pressure cannot be
measured and displayed as it is
outside the device's measuring
range.
The NIBP module is unable to
detect a pulse wave signal.
2
Remedy
If medically indicated, perform
suitable treatment or adapt the
alarm limits
(see 7.3.1, p.
Re-enter presets in the operator
menu. If this recurs: Have the
device repaired.
Attach ECG electrodes to the
patient correctly and connect the
ECG cable correctly to the ECG
electrodes
(see 4.6, p.
Have the device repaired.
Attach ECG connector of the ECG
cable to the device correctly
4.6, p.
64).
If medically indicated, perform
suitable treatment or adapt the
alarm limits
(see 7.3.1, p.
If medically indicated, perform
suitable treatment.
If medically indicated, perform
suitable treatment or adapt the
alarm limits
(see 7.3.1, p.
If medically indicated, perform
suitable treatment.
Use an alternative blood pressure
measurement device with a larger
measuring range.
Make sure that the patient lies
still when taking
measurements
Rule out the risk of patient
vibration
Use a suitable cuff size.
Ensure that the cuff is properly
attached.
Ensure that there is no
clothing between the blood
pressure cuff and the patient
111).
64).
(see
111).
111).