8. If necessary: Set alarm limits in the user menu
9. If necessary: Press the Back function button.
Result
Non-invasive blood pressure measurement has been performed.
5.8.2 Interval measurement
During an interval measurement (Interval function button), the
device performs several successive NIBP measurements. The
interval duration indicates the time between two successive NIBP
measurements.
Performing the interval measurement
Requirement
1. Press the NIBP button
2. If necessary: Select another patient group
3. Press the Interval function button.
4. Set the interval duration using the navigation knob.
Alarm settings", page
or
Set automatic alarm limits in the application menu
Application menu", page
or
Press the NIBP button
The device ends the NIBP function mode and switches to the
set mode.
An NIBP cuff is connected to the NIBP connecting tube
"4.7 Attaching the NIBP cuff", page
The device is switched on
page
76).
The patient group is set
page
78).
The device switches to NIBP function mode.
the patient group", page
The NIBP module is configured in the device accordingly with
the selected patient group.
111).
106).
for < 2 s.
68).
(see "5.1 Switching the device on",
(see "5.4 Selecting the patient group",
for < 2 s.
78).
MEDUCORE Standard
5 Operation
(see "7.3.1
(see "6
(see
(see "5.4 Selecting
2
EN
95