4.2 Connecting oxygen supply

14
EN
Assembly
5. Connect any other modules in exactly the same
way.
No more than 3 modules may be connected,
otherwise it is impossible to guarantee compliance
with the specified technical data.
6. Screw the modules and, if appropriate, the
MEDUMAT onto the assembly plate from the
back as shown in the assembly instructions
provided,
or
screw the modules to a solid wall from the back.
Please note that screws should protrude no
more than 5.5 mm into the modules.
Wash your hands thoroughly before any work on the
oxygen supply. Hydrocarbon compounds (e.g. oil,
greases, alcohols for cleaning, hand cream or
sticking plaster) can lead to explosive reactions if
they come into contact with highly-compressed
oxygen.
Under no circumstances use spanners or other
tools to tighten up the union nuts.