Pulse Oximetry Monitoring - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

15 Technical data
Sensitivity
Specificity
Precision
False positive rate
True forecast value

15.8 Pulse oximetry monitoring

Specification
SpO
display range
2
Accuracy (for all SpO
sensors which are specified in
2
the chapter entitled Scope of supply
of supply", page
190))
Pulse rate
182
EN
MEDUCORE Standard
A = Number of correct positive decisions
B = Number of false positive decisions
C = Number of false negative decisions
D = Number of correct negative decisions
This results in the following values:
Formula for the
calculation
A
------------- -
A
+
C
D
------------- -
B
+
D
A
+
D
-----------------------------------
A
+
B
+
C
+
D
B
------------- -
B
+
D
A
------------- -
A
+
B
Due to the statistical distribution of the measured values for pulse
oximetry sensors, on average two thirds of all measurements are
within the ±arm value of CO oximeters as a reference method (see
table). To determine bpm accuracy, comparative measurements
were performed with an original sensor and original monitor.
(see "16 Scope
2
Calculation
1165
----------------------- -
1165
+
31
3059
----------------------- -
11
+
3059
1165
+
3059
------------------------------------------------------ -
1165
+
11
+
31
+
3059
11
----------------------- -
11
+
3059
1165
----------------------- -
1165
+
11
Pulse oximetry
45 % to 100 %
70 % to 100 %: ≤ 2.1 %
45 % to 70 %: Not specified
20 bpm to 300 bpm ± 3 bpm
Result
97.41 %
99.64 %
99.02 %
0.36 %
99.06 %