8 Operator menu

8
Operator menu
116
EN
MEDUCORE Standard
Risk of injury due to incorrectly set parameters or too few/too
many enabled functions in the operator menu!
Incorrectly set parameters or too few/too many enabled functions
in the operator menu can result in incorrect settings in the user
menu or too limited/too comprehensive device functions. This can
cause critical operating situations and injure the patient.
⇒ The operator menu should only be used by operators who are
familiar with the settings in the operator menu and their
impacts on the user menu and device functions.
⇒ Adapt the device functions to the user's know-how.
⇒ Protect the operator menu with a password.
The operator menu contains the device presets which are
permanently stored. If the device was switched off for longer than
30 s and is switched on again, these operator menu device presets
are loaded. If the device was switched off for less than 30 s and a
patient was previously connected, the user menu settings are
restored.
Moreover, the operator menu contains functions which are only
relevant to the operator and not the user, e.g. the performance of
software updates, the enabling of optional functions and the
disabling of functions.