9 Hygienic reprocessing
Part
Device
Battery
Power supply unit and
charger
Trunk cable
Defibrillation electrodes
Pulse oximetry sensor
connecting cable
Reusable pulse oximetry
sensor
Disposable pulse oximetry
sensor
ECG cable
ECG electrodes
NIBP connecting tube
Reusable NIBP cuff
Disposable NIBP cuff
Short-term NIBP cuff
Protective transport bag
146
EN
MEDUCORE Standard
7. Carry out hygienic reprocessing of the device, components,
and accessories as specified in the table below:
Cleaning
Disinfection
Wipe with a damp
Wipe disinfection
cloth: Use water or
(Recommendation:
mild soap
terralin
Disposable item, do not reuse
Wipe with a damp
Wipe disinfection
cloth: Use water or
(Recommendation:
mild soap
terralin
Disposable item, do not reuse
Wipe with a damp
Wipe disinfection
cloth: Use water or
(Recommendation:
mild soap
terralin
Disposable item, do not reuse
Wipe with a damp
Wipe disinfection
cloth: Use water or
(Recommendation:
mild soap
terralin
Disposable item, do not reuse
Wipe disinfection
Wipe with a damp
(Recommendation:
cloth: Use water or
terralin
mild soap
maximum of 20
cycles
Wipe with a damp
Wipe disinfection
cloth: Use water or
(Recommendation:
mild soap
terralin
The applicable instructions are those in the Instructions for Use
from the manufacturers of the individual components or
accessories. Follow these Instructions for Use.
2
Thermal
disinfection
Not permitted
®
protect)
Not permitted
®
protect)
Not permitted
®
protect)
Not permitted
®
protect)
®
Not permitted
protect),
Not permitted
®
protect)
Sterilization
Not permitted
Not permitted
Not permitted
Not permitted
Not permitted
Not permitted