Weinmann MEDUCORE Standard2 Instructions For Use Manual page 85

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

After cardiac rhythm analysis and shock delivery (if necessary), the
device instructs you to carry out cardiopulmonary resuscitation.
A metronome provides a guide for chest compressions.
Voice prompt (optional)
Carry out cardiopulmonary
resuscitation
1. Perform chest compressions:
30 chest compressions for the adult patient group
15 chest compressions for the child patient group
Continuous chest compression with intubated patients
When doing so, please note: The metronome sets the ideal
frequency.
After 30/15 metronome beats, there is a pause to allow for
ventilation:
Voice prompt (optional)
Ventilate twice
2. Ventilate the patient twice.
The device outputs the message:
Voice prompt (optional)
Carry out cardiopulmonary
resuscitation
3. Repeat the cardiopulmonary resuscitation sequence.
4. If the patient is intubated: Press the Intub. function button.
The metronome sets a continuous frequency.
5. If the patients shows vital signs (breathing, response): Take
basic patient care steps.
6. After each cardiopulmonary resuscitation cycle: Check that the
defibrillation electrodes are positioned correctly.
Result
Cardiopulmonary resuscitation has been performed.
AED instruction
Cardiopulmonary resuscitation
AED instruction
Ventilate twice
AED instruction
Cardiopulmonary resuscitation
MEDUCORE Standard
5 Operation
2
EN
85