Weinmann MEDUCORE Standard2 Instructions For Use Manual page 125

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Parameter
Heart rate
Heart rate
Pulse rate
Pulse rate
SpO
SpO
Child alarm limit
settings
NIBP systolic
NIBP systolic
NIBP diastolic
NIBP diastolic
Possible
values
35/min-250/min,
in increments of 5
30/min-245/min,
in increments of 5
35/min-250/min,
in increments of 5
30/min-245/min,
in increments of 5
saturation
2
66 %-100 %
saturation
2
65 %-99 %
45 mmHg-
260 mmHg, in
increments of 5
40 mmHg-
255 mmHg, in
increments of 5
25 mmHg-
200 mmHg, in
increments of 5
20 mmHg-
195 mmHg, in
increments of 5
Description
Here you can set the upper
( ) and lower ( ) limits for
the patient group as of which
the device must emit an
alarm.
According to the values set,
the setting ranges of the
upper and lower limits
depend on each other as
follows:
The setting range for the
upper alarm limit ends at
least one setting value above
the set value for the lower
alarm limit.
The setting range for the
lower alarm limit ends at
least one setting value below
the set value for the upper
alarm limit.
MEDUCORE Standard
8 Operator menu
Factory
setting
150/min
50/min
150/min
50/min
100 %
85 %
145 mmHg
75 mmHg
100 mmHg
35 mmHg
2
EN
125