MEDUVENT Standard
Electric shock on contact with the device!
Accessories connected to the device might create a voltage on the
device. This might result in an electric shock on contact with the
device, and cause the user serious or life-threatening injury.
 Use only approved accessories.
Risk of cross-contamination resulting from use of an incorrect
patient valve!
Using a non-approved patient valve might cause rebreathing of
the previous patient into the device, contaminating the device.
This can cause serious or life-threatening injury to the patient.
 Use only patient valves approved by WEINMANN Emergency.
The device is mounted on a portable unit and is ready for use as
standard. Follow the instructions for use of the portable unit for
mounting the device on the portable unit, and mounting the
portable unit in emergency vehicles.
Failure of therapy or loss of power resulting from use of an
incorrect power supply unit and charger!
If you are using a portable unit with the MEDUVENT Standard and
MEDUCORE Standard or MEDUCORE Standard
combination, the devices might lose power when using the 50 W
power supply unit and charger, and therapy might fail
prematurely.
 Use only the more powerful 100 W power supply unit and
charger (WM 28937) for the MEDUVENT Standard and
MEDUCORE Standard/MEDUCORE Standard
combination.
2
device
2
device