8.2.1 Menu structure
 
  
! 
 
 
  
  
#   
"  
8.2.2 Possible values
Parameter
Adult AED settings/child AED settings
First shock
Energy curve
Further shocks
CPR phase
Duration
AED settings
for adults
AED settings
Child
Pictograms in
parameter view
Charging during
analysis
Possible
Description
values
Here you can set the shock
energy of the first shock in
the AED mode.
1 J to 200 J
Here you can set the shock
energy of subsequent shocks
in the AED mode.
Here you can set the interval
120 s to 300 s
between the cardiac rhythm
analyses.
Energy curve
CPR phase
Audio outputs
Automatic
start of analysis
MEDUCORE Standard
8 Operator menu
First shock
Further shocks
Duration
Ventilation pause
CV ratio
Start in intub. mode
CPR
voice prompt
Metronome
Metronome frequency
Factory
setting
AED mode
adult: 150 J
AED mode
child: 75 J
AED mode
adult: 200 J
AED mode
child: 75 J
120 s
2
EN
119