Defibrillation electrode packaging
ECG electrode packaging
ECG cable
Pulse oximetry sensor connecting cable
Pulse oximetry sensor
NIBP connecting tube
NIBP cuff
2. If necessary: Replace parts.
3. Check the expiry date on the ECG electrodes and defibrillation
electrodes.
If necessary: Replace ECG electrodes and/or defibrillation
electrodes.
4. Switch on the device
76).
5. Select the menu item Function check in the start menu.
The automatic function check starts and the date of the last
successful function check is displayed.
6. Connect the function test resistor to the trunk cable.
(see "5.1 Switching the device on", page