5 Operation
Result
88
EN
MEDUCORE Standard
Risk of injury from electric shock!
The electric shock administered to the patient may result in injury
to the patient, user or bystanders.
⇒ Stand clear of the patient.
⇒ Keep patient away from liquids (e.g. blood, gel or saline
solution).
⇒ Do not touch parts in contact with the patient (e.g. bed frames
or stretchers).
⇒ Keep your distance from liquids in contact with the patient.
⇒ Clearly warn bystanders to stand clear of the patient or parts in
contact with the patient and to keep their distance from liquids
in contact with the patient.
Damage to the device caused by removal of the defibrillation
electrodes during shock delivery!
Removal of the defibrillation electrodes during shock delivery can
cause damage to the device.
⇒ Always leave defibrillation electrodes connected to the device
and the patient during shock delivery.
8. Deliver a shock with the shock button
When doing so, please note: If the shock button
pressed, the capacitor for the shock energy discharges
automatically after 30 seconds.
9. If necessary: Cancel shock charging by pressing the Cancel
function button or by switching to another mode.
The patient receives an electric shock.