11 Alarms and faults
Alarm
Plug in pad connector
(Monitor mode)
Pulse rate
Pulse rate
Memory error
SpO
2
SpO
2
SpO
module defective
2
NIBP systolic
NIBP systolic
162
EN
MEDUCORE Standard
Cause
Pad connector of the defibrillation
electrodes not connected or not
correctly connected to the trunk
cable, or trunk cable not
connected or not correctly
connected to the device.
Measured pulse rate is above the
set alarm limit.
Measured pulse rate is below the
set alarm limit.
Internal memory module defective. Have the device repaired.
Measured oxygen saturation is
above the set alarm limit and the
SpO
signal quality is ≥ 40 %.
2
Measured oxygen saturation is
below the set alarm limit and the
SpO
signal quality is ≥ 40 %.
2
Internal SpO
module defective.
2
Measured systolic blood pressure
is above the set alarm limit.
Measured systolic blood pressure
is below the set alarm limit.
2
Remedy
Connect the pad connector of the
defibrillation electrodes correctly
to the trunk cable, or connect the
trunk cable correctly to the device.
If medically indicated, perform
suitable treatment or adapt the
alarm limits
(see 7.3.1, p.
If medically indicated, perform
suitable treatment or adapt the
alarm limits
(see 7.3.1, p.
If medically indicated, perform
suitable treatment or adapt the
alarm limits
(see 7.3.1, p.
If medically indicated, perform
suitable treatment or adapt the
alarm limits
(see 7.3.1, p.
Have the device repaired.
If medically indicated, perform
suitable treatment or adapt the
alarm limits
(see 7.3.1, p.
If medically indicated, perform
suitable treatment or adapt the
alarm limits
(see 7.3.1, p.
111).
111).
111).
111).
111).
111).