Settings - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

6 Application menu

6.3 Settings

Possible
Parameter
values
25 %
50 %
Volume
75 %
100 %
for adults
Child
Patient group
Infant
Autom. alarm limits
Activated
Night colors
Deactivated
108
EN
MEDUCORE Standard
Description
Here you can set the volume of the device for
the current session.
Here you can select the patient group.
The device sets the alarm limits for the
physiological alarms automatically. The alarm
limit corridor is 10 %, 20 % or 30 % above
or below the physiological values measured
at the time of activation. The alarm limit
corridor can be set in the user menu
User menu", page
109).
Here you can set whether the device is to
display night colors.
2
Factory setting
75 %
For adults
20 %
(see "7
Deactivated