11 Alarms and faults

11.2 Alarm messages

11.2.1 High-priority alarm (red)
Alarm
Battery defective
Insert battery
Battery empty
158
EN
MEDUCORE Standard
Type of alarm
High priority
signal
Alarm light
Color
Red
Frequency
2 Hz
Duty cycle
33 % on
Size
10 mm x 35 mm
The device additionally displays physiological alarms through the
flashing of the respective parameter field.
The device's alarm behavior depends on the mode selected:
In AED mode, the device does not emit alarms. The device
provides information which could result in the AED mode being
impaired via AED statement texts and voice prompts.
The audio alarm output is deactivated in manual mode. It can
be reactivated at any time by pressing the alarm button. In the
operator menu, the operator can specify whether the audio
alarm output is to be activated upon activating manual mode.
Cause
Battery defective.
Battery not inserted or incorrectly
inserted (manual mode).
Very low battery status.
2
Medium
Low priority
priority
Off
Off
Not applicable
Not applicable
Not applicable
Not applicable
Remedy
Let the device run on battery
power without line power until it
switches off. Fully recharge battery
(see 4.3.2, p.
55). If the device
continues to display the alarm:
Replace battery
(see 4.3.4, p.
Insert battery correctly
53).
Replace battery
(see 4.3.4, p.
Connect device to the line power
(see 4.2, p. 53)
and charge battery
(see 4.3.2, p.
55).
56).
(see 4.2, p.
56).