Performing the NIBP measurement
Requirement
Risk of injury due to incorrectly selected patient group
The device only provides correct measured values if the
appropriate patient group is selected. An incorrect patient group
can lead to incorrect measurements and patient injuries.
⇒ Select the suitable patient group for the patient
Risk of injury due to incorrectly selected NIBP cuff
The device only provides correct measured values if the
appropriate NIBP cuff is selected. An unsuitable NIBP cuff can lead
to incorrect measurements and patient injuries.
⇒ Select the suitable NIBP cuff for the patient
1. Press the NIBP button
2. If necessary: Select another patient group
Patient position (ideal position: Sitting comfortably, legs not
crossed, feet flat on the floor, back and arm supported, middle
of the NIBP cuff level with the right heart atrium)
Exertion (recommendation: Patient should rest for 5 minutes
before the measurement, keep still during the measurement
and not speak)
Physiological condition.
An NIBP cuff is connected to the NIBP connecting tube
"4.7 Attaching the NIBP cuff", page
The device is switched on
page
76).
The patient group is set
page
78).
Selecting the patient group", page
the NIBP cuff", page
The device switches to NIBP function mode.
the patient group", page
The NIBP module is configured in the device accordingly with
the selected patient group.
68).
(see "5.1 Switching the device on",
(see "5.4 Selecting the patient group",
78).
68).
for < 2 s.
78).