11 Alarms and faults
Fault
Device failure
Alarm LED flashes
Audio alarm output
11.3.2 Defibrillation
Fault
The Check pad electrodes
alarm appears although the
defibrillation electrodes are
attached to the patient and to the
device via the trunk cable.
11.3.3 Battery
Fault
Red fault indicator lights up when
status button on battery is pressed
or red battery status indicator on
device lights up
Battery does not respond when
status button is pressed
Device runtime with battery
operation too short
166
EN
MEDUCORE Standard
Cause
Device defective.
Cause
Defibrillation electrodes incorrectly
attached to the patient or pad
connector incorrectly connected to
the trunk cable or the trunk cable
is incorrectly connected to the
device.
Cause
Battery defective.
Battery temperature outside the
permitted range (> 70 °C).
Battery has run down completely
and has shut down to prevent
deep discharge.
Battery has reached end of its
service life.
2
Remedy
Have the device repaired.
Remedy
Attach defibrillation electrodes to
the patient correctly
(see 4.4, p.
57).
Connect pad connector of the
defibrillation electrodes to the
device correctly.
Replace defibrillation electrodes.
Connect the trunk cable correctly
to the device. If necessary: Replace
the trunk cable.
Have the device repaired.
Remedy
Replace battery.
Use battery within permitted
temperature range
(see 15.3, p.
178).
Charge battery in the device for
24 hours
(see 4.3.2, p.
24 hours:
Green LED is lit: Battery fully
charged and ready for use.
Red LED or no LED is lit: Battery
defective. Replace battery.
Replace battery.
55). After