8 Operator menu
Parameter
Adult AED settings/child AED settings
Ventilation pause
CPR phase
CV ratio
Start in intub.
mode
120
EN
MEDUCORE Standard
Possible
Description
values
Here you can set the time
interval for ventilation
between the chest
compressions.
When choosing a setting,
consider whether the voice
prompt is deactivated:
3 s to
8 s
Here you can set the
compression/ventilation
ratio.
30:2
15:2
Activated
Here you can set whether the
device is to start up with the
continuous metronome
Deactivated
rhythm for intubated
patients.
2
When the voice prompt
is activated, the pause
begins after the last
metronome sound and
ends when the voice
prompt begins.
When the voice prompt
is deactivated, the pause
begins after the last
metronome sound and
ends when the first
metronome sound
begins.
15:2: After
15 metronome beats
there is a pause to allow
for ventilation.
30:2: After
30 metronome beats
there is a pause to allow
for ventilation.
Factory
setting
5 s
AED mode
adult: 30:2
AED mode
child: 15:2
Deactivated